Exclusively available in india the DP-60 Dentalness Air Polisher can be used for air-polishing scaling, Remove dental plaque, tartar and other tooth stains on the tooth surface plaque from periodontal tissues, and polish the tooth surface.
DP-60 Dentalness Air Polisher can be used for air-polishing scaling.It has the following features:
The three levels meet different clinical needs, and the humanized design improves patient comfort.
Product Structure and Composition :
It consists of function control circuit, powder tank, air circuit, liquid circuit,handpiece, air polishing nozzle, foot switch (wired ) and power supply
Product Application Scope: Remove dental plaque, tartar and other tooth stains on the tooth surface, removetartar and plaque from periodontal tissues, and polish the tooth surface
Product Working Principle: Compressed air is used to deliver dental powder particles such as sodium bicarbonate from the powder tank to the tooth surface, to remove residual tartar and pigments on the teeth, thereby achieving the effect of polishing the tooth surface
Contraindications
Parameter | Specification |
---|---|
Control | Touch Panel |
Handpiece | Sandblasting Handpiece AH-1 ×1 |
Operating Mode | A/C |
Water Tank | 800 mL ×1 |
External Water Hose | Unavailable |
Foot Pedal | Wired (IPX1) ×1 |
Powder Tank Components | 1 |
Small Cart | Optional |
Weight | 4.8 kg |
Input Voltage | 100–240 V ~ 50/60 Hz |
Input Power | 20 VA |
Water Pressure | 0.01 MPa–0.5 MPa (0.1–5 bar) |
Compressed Air Supply | 5–6 bar (0.5–0.6 MPa) |
Packing Size | 415 × 175 × 355 mm |
Storage Environment | Temp: –20 °C to 55 °C |
Relative Humidity | ≤ 90 % |
Atmospheric Pressure | 70–106 kPa |
Working Environment | Temp: 5–40 °C Humidity: ≤ 80 % Pressure: 70–106 kPa |
Lifespan | 10 years |
Handpiece Sterilization | 134 °C and 0.22 MPa pressure |
No. | Item | Qty | Model/Spec |
---|---|---|---|
1 | Main Unit | 1 | DP-60 |
2 | Power Cord | 1 | |
3 | Wired Foot Switch | 1 | |
4 | Air Polishing Handpiece | 1 | AH-1 |
5 | Water Bottle | 1 | |
6 | Needle | 1 | 0.5 |
7 | Needle | 1 | 1 |
8 | Powder Tank (Supragingival) | 1 | |
9 | Hose Assembly | 1 | |
10 | T Type Quick Connector | 2 | PEG8-6 |
11 | O-ring | 6 | 11.1×1.8 |
12 | O-ring | 2 | 5×1 |
13 | O-ring | 1 | 43×2 |
14 | O-ring | 6 | 8×1.5 |
15 | O-ring | 6 | 6.5×1.5 |
16 | O-ring | 6 | 11×1.5 |
17 | O-ring | 6 | 4.5×1.5 |
18 | Silicone Tube | 2 | 4×1.5 / 11mm |
19 | Instruction Manual | 1 |
To ensure optimal performance and long-term use of the air polishing unit, and to enhance work efficiency and operational safety, the device should be maintained according to the following guidelines:
I. Daily Maintenance
1.1 After removing the disposable protective blue film, promptly wipe the surface of the device with a damp cloth to keep it clean and prevent cross-contamination.
1.2 Perform a one-touch rinse before and after each sandblasting session to ensure the tubing is functioning properly and to flush out any remaining powder.
1.3 After use, pour out any remaining powder in the powder tank. Use an air gun to clean residual powder from the tank lid, tank threads, and inside the tank.
1.4 Rinse the air polishing handpiece and its connecting cable with water to remove residual powder. After rinsing, dry thoroughly to avoid jamming or clogging.
II. Periodic Maintenance
2.1 The air polishing handpiece must be cleaned regularly. Use an air gun, sand flow cleaner, or sand flow needle for unclogging. If clogging is severe, soak the handpiece in hot water for 15 minutes before clearing.
2.2 Clean the water tank surface regularly to prevent limescale buildup.
2.3 Regularly check and clean the water-gas separator valve at the rear of the device to ensure the air source remains dry.
III. Inspection and Replacement of Accessories
3.1 Shake the powder evenly before each use to ensure the particles are uniform and dry before pouring into the tank.
3.2 Before each use, confirm that air pressure, powder, powder flow, and water output are all properly adjusted and functioning.
3.3 Periodically check connectors on the tail line, powder tank cap, water tank, and air source fittings. Ensure that the sealing rings at these positions are intact. Replace any worn or damaged parts in time to maintain polishing and cleaning performance.
3.4 Before activating the water heating function, make sure the water heating switch in the settings is turned ON.
3.5 Before using the wireless foot pedal, confirm that two AA batteries are installed in the pedal and that the wireless pedal function is turned ON in the settings.
IV. Protection and Storage
4.1 Store the device and consumables in a dry and well-ventilated area to prevent moisture and corrosion. Protect the main unit, handpiece, and water tank from impact or damage.
4.2 The device should be flushed with water at least once a month for 5 minutes each time to maintain internal cleanliness.
1 year in main unit & 6months for handpiece